SW01-R4 RS485 Point Water Leak Sensor
HANS SW01-R4 is an RS485 point-type water-leak alarm transmitter for detecting conductive liquid leakage using a point probe or compatible leak-detection rope. It uses a high-resolution sampling circuit, interference-filtering circuitry and a low-power MCU for liquid-leak detection. The device communicates through RS485 using Modbus-RTU; a dry-contact output is not confirmed by the V1.1 technical table. It is externally powered by 12–24 V DC and supports configurable baud rates and device addressing. Typical applications include basements, equipment rooms, hotels, libraries, warehouses, telecommunications sites, kitchens, bathrooms and balconies where leakage or overflow may occur.


PRODUCT TYPE
RS485 Point Water Leak Sensor
TECHNOLOGY
TECHNICAL DATA
Technical Specifications — HANS SW01-R4
| Model | SW01-R4 (manufacturer source model: SD123-SW01-R4) |
|---|---|
| Product type | RS485 point-type water-leak alarm sensor |
| Detection principle | Conductive liquid detection using point probe / leak-detection rope |
| Communication interface | RS485 |
| Communication protocol | Modbus-RTU |
| Digital contact output | Normally-open (NO) contact |
| Power supply | 12–24 V DC, external supply |
| Baud rate | 1,200 / 2,400 / 4,800 / 9,600 / 57,600 / 115,200 bit/s; configurable; factory default 9,600 bit/s |
| Device address | 1–252, configurable; factory default 1 |
| Data format | 8 data bits, no parity, 1 stop bit (8N1) |
| Error checking | 16-bit CRC |
| Function code | Modbus function code 0x03 for register reading |
| Ingress protection | IP65 |
| Operating temperature | 0 °C to +50 °C |
| Storage temperature | −20 °C to +60 °C |
| Operating humidity | 0–95% RH |
| Storage humidity | 0–95% RH |
| Pressure condition | Standard atmospheric pressure (100 kPa) ±10% |
| Detected liquids | Tap water and groundwater |
| Installation | Wall-mounted enclosure; leak-detection rope may be positioned vertically or laid along multiple monitoring points / wall edges |
Wiring: Red = DC positive; black = DC negative; yellow = RS485 A; white = RS485 B.
Important installation note: Do not place the sensor enclosure where it can be submerged. Avoid highly corrosive atmospheres (including H₂S, SOx, Cl₂ and HCl) and strong magnetic fields. Protect the unit from impact and vibration.
Source: Manufacturer manual SD123-SW01-R4 V1.1, revised 2025-09-25, and official product page 7569339. The source page and manual specify an NO contact; no relay rating, probe cable length or alarm delay is inferred.
SPECIFICATIONS
Technical Specifications — HANS SW01-R4
| Model | SW01-R4 (manufacturer source model: SD123-SW01-R4) |
|---|---|
| Product type | RS485 point-type water-leak alarm sensor |
| Detection principle | Conductive liquid detection using point probe / leak-detection rope |
| Communication interface | RS485 |
| Communication protocol | Modbus-RTU |
| Digital contact output | Normally-open (NO) contact |
| Power supply | 12–24 V DC, external supply |
| Baud rate | 1,200 / 2,400 / 4,800 / 9,600 / 57,600 / 115,200 bit/s; configurable; factory default 9,600 bit/s |
| Device address | 1–252, configurable; factory default 1 |
| Data format | 8 data bits, no parity, 1 stop bit (8N1) |
| Error checking | 16-bit CRC |
| Function code | Modbus function code 0x03 for register reading |
| Ingress protection | IP65 |
| Operating temperature | 0 °C to +50 °C |
| Storage temperature | −20 °C to +60 °C |
| Operating humidity | 0–95% RH |
| Storage humidity | 0–95% RH |
| Pressure condition | Standard atmospheric pressure (100 kPa) ±10% |
| Detected liquids | Tap water and groundwater |
| Installation | Wall-mounted enclosure; leak-detection rope may be positioned vertically or laid along multiple monitoring points / wall edges |
Wiring: Red = DC positive; black = DC negative; yellow = RS485 A; white = RS485 B.
Important installation note: Do not place the sensor enclosure where it can be submerged. Avoid highly corrosive atmospheres (including H₂S, SOx, Cl₂ and HCl) and strong magnetic fields. Protect the unit from impact and vibration.
Source: Manufacturer manual SD123-SW01-R4 V1.1, revised 2025-09-25, and official product page 7569339. The source page and manual specify an NO contact; no relay rating, probe cable length or alarm delay is inferred.
KEY FEATURES
- Point-type liquid-leak detection for tap water and groundwater.
- RS485 interface with Modbus-RTU protocol.
- Normally-open (NO) contact output.
- External 12–24 V DC power supply.
- Six configurable baud rates from 1,200 to 115,200 bit/s.
- Configurable device address from 1 to 252.
- IP65 enclosure protection.
- Wall-mounted installation with point probe or leak-detection rope.
PRODUCT SYSTEM